Obsolete and Discontinued Commands 7
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 745
:MTESt:{STARt | STOP}
(see page 798)
Command Syntax
:MTESt:{STARt | STOP}
The :MTESt:{STARt | STOP} command starts or stops the acquisition
system.
See Also "Introduction to :MTESt Commands" on page 372
NOTE
The :MTESt:STARt and :MTESt:STOP commands are obsolete and are provided for
backward compatibility to previous oscilloscopes. Use the :RUN command (see page 186)
and :STOP command (see page 190) instead.