Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 375
:MTESt:AMASk:CREate
(see page 798)
Command Syntax
:MTESt:AMASk:CREate
The :MTESt:AMASk:CREate command automatically constructs a mask
around the current selected channel, using the tolerance parameters
defined by the :MTESt:AMASk:XDELta, :MTESt:AMASk:YDELta, and
:MTESt:AMASk:UNITs commands. The mask only encompasses the portion
of the waveform visible on the display, so you must ensure that the
waveform is acquired and displayed consistently to obtain repeatable
results.
The :MTESt:SOURce command selects the channel and should be set before
using this command.
See Also "Introduction to :MTESt Commands" on page 372
":MTESt:AMASk:XDELta" on page 378
":MTESt:AMASk:YDELta" on page 379
":MTESt:AMASk:UNITs" on page 377
":MTESt:AMASk:SOURce" on page 376
":MTESt:SOURce" on page 402
Example Code "Example Code" on page 372