Agilent Technologies InfiniiVision 6000 Series Manual de servicio Pagina 572

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 934
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 571
572 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:LIN:SAMPlepoint
(see page 798)
Command Syntax
:TRIGger:LIN:SAMPlepoint <value>
<value><NL>
<value> ::= {60 | 62.5 | 68 | 70 | 75 | 80 | 87.5} in NR3 format
The :TRIGger:LIN:SAMPlepoint command sets the point during the bit time
where the bit level is sampled to determine whether the bit is dominant or
recessive. The sample point represents the percentage of time between the
beginning of the bit time to the end of the bit time.
Query Syntax
:TRIGger:LIN:SAMPlepoint?
The :TRIGger:LIN:SAMPlepoint? query returns the current LIN sample
point setting.
Return Format
<value><NL>
<value> ::= {60 | 62.5 | 68 | 70 | 75 | 80 | 87.5} in NR3 format
See Also "Introduction to :TRIGger Commands" on page 479
":TRIGger:MODE" on page 485
":TRIGger:LIN:TRIGger" on page 577
NOTE
The sample point values are not limited by the baud rate.
Vista de pagina 571
1 2 ... 567 568 569 570 571 572 573 574 575 576 577 ... 933 934

Comentarios a estos manuales

Sin comentarios