474 Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TIMebase:SCALe
(see page 798)
Command Syntax
:TIMebase:SCALe <scale_value>
<scale_value> ::= 500 ps through 50 s in NR3 format
The :TIMebase:SCALe command sets the horizontal scale or units per
division for the main window.
Query Syntax
:TIMebase:SCALe?
The :TIMebase:SCALe? query returns the current horizontal scale setting in
seconds per division for the main window.
Return Format
<scale_value><NL>
<scale_value> ::= 500 ps through 50 s in NR3 format
See Also "Introduction to :TIMebase Commands" on page 468
":TIMebase:RANGe" on page 471
":TIMebase:WINDow:SCALe" on page 478
":TIMebase:WINDow:RANGe" on page 477