Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 471
:TIMebase:RANGe
(see page 798)
Command Syntax
:TIMebase:RANGe <range_value>
<range_value> ::= 5 ns through 500 s in NR3 format
The :TIMebase:RANGe command sets the full- scale horizontal time in
seconds for the main window. The range is 10 times the current
time- per-division setting.
Query Syntax
:TIMebase:RANGe?
The :TIMebase:RANGe query returns the current full- scale range value for
the main window.
Return Format
<range_value><NL>
<range_value> ::= 5 ns through 500 s in NR3 format
See Also "Introduction to :TIMebase Commands" on page 468
":TIMebase:MODE" on page 469
":TIMebase:SCALe" on page 474
":TIMebase:WINDow:RANGe" on page 477
Example Code
' TIME_RANGE - Sets the full scale horizontal time in seconds. The
' range value is 10 times the time per division.
myScope.WriteString ":TIM:RANG 2e-3" ' Set the time range to 0.002
seconds.
Example program from the start: "VISA COM Example in Visual Basic" on
page 824