4 Operating the 8000
Daisy Chaining
A series of dissolution apparatus and 8000s can be started at the same
time using one 8000 as the master unit. To set up a daisy chain,
complete the following steps:
1 Connect the master 8000 START OUTPUT jack to the START
INPUT jack of its dedicated dissolution apparatus using a four-pin
cable. Connect the START OUTPUT jack of the first dissolution
apparatus to the START INPUT jack of a second 8000. Connect the
START OUTPUT jack of the second 8000 to the START INPUT jack
of the second dissolution apparatus. Daisy chain successive units
using the same technique.
2 Set the desired program for each 8000 independently.
3 For each of the remote units:
• Press START PROG.
• Enter the desired program number and press ENTER.
• Select option 1, Run Single Prog.
• Select option 3, Remote Start.
• Select option 1, Load & Run DISSO or APP 3/7. The program
loads and the Ready screen displays.
4 For the master 8000:
• Press START PROG.
• Enter the desired program number and press ENTER.
• Select option 1, Run Single Prog.
• Select option 1, Start Now.
• Select option 1, Load & Run DISSO or APP 3/7. The program
starts.
Ensure the communication port identification number for each linked
8000 increases incrementally (see “4 Set Com Port ID” under “Printer
Operation and Communications” on page 103.