86 8000 Operator’s Manual
4 Operating the 8000
Creating a Sample Program
1 From the Ready screen, press PROG to program the 8000.
2 Enter a program number and press ENTER. The Program Variables
screen displays, as shown below.
5 Filter Changer Use this option to manually operate the filter changer.
Select option 5, Filter Changer. The Filter Change Manual Control screen displays.
To open the filter clamp, select option 1, Open Clamp.
To close the filter clamp, select option 2, Close Clamp.
To change the filters, select option 3, Change Filters. Enter the number of filters to
change and press ENTER. The specified number of filters are expelled and
replaced.
Press ESC to return to the Operate Syringe/Filter Changer screen.
6 Syringe Parameters Use this option to configure the syringe pump. Select option 6, Syringe Parameters.
The Set Syringe Pump Parameters screen displays. For a description of the screen
options, see “Main Menu Option 3 - Set Syringe Pump” on page 55.
Note: The syringe pump parameters are global. These parameters can be entered
under option 3 of the Main Menu or option 6 of the Operate Syringe/Filter Changer
screen.
NOTE
Programs are not write-protected. Check with other users before
proceeding.
Table 2 1 Operate Syringe/Filter Changer Options
Option Function
NOTE
The 8000 can store up to 15 programs in non-volatile memory. Valid
program numbers are 1 to 15. If 0 (zero) or a number larger than 15 is
entered, the system continues to prompt you until a valid number is
entered.