108 8000 Operator’s Manual
5 Maintenance and Troubleshooting
Start Immediately
From the Cleaning Systems Selections screen, select option 1, Start
Clean. The 8000 immediately starts to clean the system by priming the
lines. Once the system completes the priming, the needle manifold jogs
up and down while opening and closing the valves. This motion rinses
the needles in the 8000 rinse chamber. Then the system purges the
cleaning solution from the sample lines.
The following figure is a generalized flow diagram of the clean cycle:
Figure 16 The Cleaning Process
Programmed Clean
To run a program concurrently with the sampling program which rinses
the needles in a predetermined cleaning solution contained in the rinse
chamber, complete the following steps:
1 Ensure the rinse chamber of the media rinse reservoir is filled with
the cleaning solution.
2 From the Cleaning Systems Selections screen, select option 2,
Enable Auto Clean. Option 3, Disable Auto Clean, displays on the
Cleaning Systems Selections screen and option 2 changes to Program
Place Probes
in Rinse
Sol ution
Prime Tubing
Move
Head to Rinse
Position
Open/Close
Valves
Reverse Pump
Open/Close
Valves
Raise/Lower
Needles
Purge Tubing
(Clear Filters)
Start
Done