Maintenance and Troubleshooting 5
8000 Operator’s Manual 115
System Cleaning Procedure with Syringe Pump and
Filter Changer (optional)
To clean the 8000, syringe pump, and filter changer system, use the
appropriate cleaning solution based on the product being tested. Ensure
the sampling cannulas are submerged into the cleaning solution.
Complete the following steps to set up the cleaning procedure:
1 From the Ready screen, press PROG to program the 8000.
2 Enter a program number (1 - 15) to use for this cleaning program and
press ENTER. The Program Variables screen displays.
3 Select option 1, Set DISSO Parameters.
4 Enter the following parameters:
5 Press ESC > Y. The Program Variables screen displays.
NOTE
Ensure the RM option is disabled by pressing MENU > 0 > R.
Table 2 9 Disso Parameters.
Selection Response
1 Set Spindle RPM Enter 50 and press ENTER.
2 Final Spin Length Enter 0 and press ENTER.
3 Ini/Final Vessel Temp Select option 2, Disable, for each selection.
4 Set Bath Temp Enter 37.3 and press ENTER.
Note: It is not necessary to set the water bath temperature if using a dissolution
apparatus with Direct Vessel Heating (DVH).
5 Com Port ID Enter 01 and press ENTER.
6 DDM or Baskets Select option 2, Paddles. Select option 2, Simultaneous.
7 Print Freq Enter 000 and press ENTER.