Maintenance and Troubleshooting 5
8000 Operator’s Manual 107
5 On the 8000, press CLEAN SYSTEM. The Cleaning Systems
Selections screen displays.
To start the cleaning process immediately, see “Start Immediately” on
page 108. To program the auto clean function, see “Programmed
Clean” on page 108.
Apparatus 3 / 7
1 Place the appropriate number of outer media tubes in Row 1 of the
Apparatus 3 / 7.
2 Fill each outer media tube with an appropriate cleaning solution,
usually water.
3 From the Main Menu, select option 6, Control APP 3/7. See “6
CONTROL APP 3/7” on page 74.
4 Select option 3, Goto Row.
5 Enter 1 and press ENTER to move the drive unit to Row 1 which
contains the outer media tubes filled with cleaning solution.
6 Select option 1, Dip, to lower the agitator shafts and begin dipping. A
slow dipping speed, such as 5 DPM, is recommended. See “1
DIPPING SPEED (DPM)” on page 79 for information about how to
set the dipping speed.
7 Once dipping has started, press CLEAN SYSTEM on the 8000. The
Cleaning Systems Selections screen displays.
To start the cleaning process immediately, see “Start Immediately” on
page 108. To program the auto clean function, see “Programmed
Clean” on page 108.
***CLEANING SYSTEMS SELECTIONS***
1 START CLEAN 2 ENABLE AUTO CLEAN
TIME: 01/01/12 08:28:46 ELAPSE: 000:00:00
***CLEANING SYSTEMS SELECTIONS***
1 START CLEAN 2 ENABLE AUTO CLEAN
TIME: 01/01/12 08:28:46 ELAPSE: 000:00:00