Operating the 8000 4
8000 Operator’s Manual 67
3 To run a single program, select option 1.
To link another program to the one previously selected, select option 2.
The following screen displays:
Enter the program number to link to the current program and press
ENTER.
4 The Select Start Mode screen displays. Options 1 - 3 start the
collection program. Option 4 allows you to change or review the
program before beginning.
If more than ten rows are required to collect samples, linking a program
allows you to run two programs with identical test parameters back to
back. To link identical programs, link the program to itself.
LINK PROG #: (1-15):__
TIME: 01/01/12 15:02:30 ELAPSE: 000:00:04
When running two programs, it is necessary to replace the full sample
tray with an empty one between programs. Enable the program
completion alarm (see “5 Prg Compl Alarm” under “Main Menu Option
1 Set Clock/Alarms” on 40) to alert the analyst when the first program
completes and the second program begins.
***SELECT START MODE***
1 START NOW 2 DELAYED START
3 REMOTE START 4 MODIFY PROGRAM
PROG#: 1 R-M: DISABLED VOL: 5 ml