70 8000 Operator’s Manual
4 Operating the 8000
Select DISSO Program Start Mode Screen
Following is a description of the Select DISSO Program Start Mode
screen options:
3 Run 8000
Only
Select option 3, Run 8000 Only, to
immediately start the 8000 only. The
8000 executes the program currently
selected.
This option is used when other
instrumentation, such as a dissolution
apparatus, is connected to the 8000
through the START OUTPUT jack on
the rear of the instrument.
Select option 3, Run 8000 Only, to start the 8000 only.
Enter the start time in 24-hour format and press ENTER.
Enter the start date in the appropriate format and press
ENTER. The DISSO Options screen displays.
If the delayed start alarm is enabled, enter the duration
before the start of the program you would like the alarm to
sound. To disable the alarm, see MENU > 0 > C under
“Hidden Key Functions” on page 41.
This option is used when other instrumentation, such as a
dissolution apparatus, is connected to the 8000 through the
START OUTPUT jack on the rear of the instrument.
4 Set DISSO Use this option to modify or review
the current program parameters. Select
option 4, Set DISSO. The Set DISSO
Parameters screen displays. See “Main
Menu Option 2 - Set DISSO” on
page 53 for a description of the DISSO
Parameters screen options.
Table 1 2 Disso Options (Apparatus 1 / 2)
Option Start Now Function Delayed Start Function
***SELECT DISSO PROGRAM START MODE***
1 INSTANT 2 SET VESSEL TEMP AND START
TIME: 01/01/12 08:09:51 ELAPSE: 000:00:00
Vista de pagina 69
1 2 ... 65 66 67 68 69 70 71 72 73 74 75 ... 125 126

Comentarios a estos manuales

Sin comentarios