Setting Up the 8000 3
8000 Operator’s Manual 33
starting with Position 1.
Figure 11 Resident Probe Housing and Sample Cannula
13 Connect a separate length of tubing from the right port (labeled
OUT) of each syringe to the top inlet ports of the filter changer
starting with the port closest to the carousel.
14 Connect another separate length of tubing lines from the lower side
ports of the filter changer to the tubing coming from the left positions
of the 8000 solenoid valves. Ensure each line is attached to its
identical color.
15 Connect the tubing coming from the top positions of the 8000
solenoid valves to the return lines of the dissolution apparatus.
Ensure each line is attached to its identical color.
16 If the system configuration includes the 705-DS, connect the return
lines to the return cannulas of the resident probes.
17 To install the catch basin for the filters, slide the side of the basin
containing the opening over the screw mounted on the front of the
filter changer.
sample cannula
resident probe housing
WARNING
Do not tighten the plastic fittings on the ends of the tubing more
than half a turn past finger-tight. Overtightening will damage
the fittings.