52 8000 Operator’s Manual
4 Operating the 8000
4 Calib. Compl Alarm Use this option to set an alarm to sound when autocalibration is complete and all
lines are purged.
Select option 4, Calib. Compl Alarm. Select option 1 to enable or option 2 to
disable the alarm. The Set Clock and Alarms screen displays.
Press any key to silence the alarm after it sounds.
Note: This option is only available if your system configuration includes the
peristaltic pump.
5 Prg Compl Alarm Use this option to set an alarm to sound at the end of the program. It is a useful
reminder that the program is complete.
Select option 5, Prg Compl Alarm. Select option 1 to enable or option 2 to disable
the alarm. The Set Clock and Alarms screen displays.
Press any key to silence the alarm after it sounds.
6 Start Output Delay Note: Option 6 displays only when the syringe pump is enabled.
If the syringe pump is enabled, use this option to set the delay from the start of a
program until the TTL signal is sent from the START OUTPUT jack on the 8000
rear panel.
Select option 6, Start Output Delay. The Start Output Delay screen displays. Enter
a delay length of up to 99:59 minutes for the 8000 TTL output and press ENTER.
The Set Clock and Alarms screen displays.
Note: This output signal can be used to start other instrumentation, such as a
dissolution apparatus, after the program on the 8000 is started. The default delay is
00:00, indicating that the signal will be sent at the moment a program starts. For
example, a value of 10:00 delays the signal for ten minutes after the program
starts. Set the value to 00:00 to eliminate the delay feature.
When the syringe pump is disabled, this option displays under the Main Menu.
See “3 Start Output Delay (peristaltic pump)” on page 49.
Table 4 Set Clock and Alarms Options (Apparatus 1 / 2)
Option Function