40 8000 Operator’s Manual
3 Setting Up the 8000
Loading Filters (Filter Changer only)
1 Choose a tube of filters. Although you can load up to eight tubes of
filters, only one tube is necessary for the filter changer to function
properly.
2 Remove the cap from the end of the tube to expose the long, thin exit
nozzle of the end filter.
3 Insert the tube, with the open end down, into the metal filter
assembly in the filter changer carousel. Pivot the tube until it slides
vertically into the white stabilizer slot.
4 Repeat Steps 2 - 4 to load additional tubes of filters.
Changing Filter Types (Filter Changer Only)
1 Remove all tubes of filters from the carousel of the filter changer.
2 From the 8000 Ready screen, press MENU. The Main Menu
displays.
3 Select option 7, Manual Operation. The Manual Operations screen
displays.
4 Select option 6, Syringe/Filter.
5 From the Operate Syringe/Filter Changer screen, select option 5,
Filter Changer. The Filter Changer Manual Control screen displays.
6 Select option 3, Change Filters.
7 Enter 8 as the number of filters to change and press ENTER. The
filters expel from the filter changer.
8 Repeat Steps 6 and 7 until filters no longer expel from the filter
changer. The following error message displays on the screen:
FILTER
CHANGER ROTOR ERROR. THE FILTER CHANGER IS EMPTY OR HAS
FAILED TO LOAD PROPERLY. PRESS <1> TO CONTINUE OR <ESC> TO
QUIT.
9 Press ESC. The Filter Changer Manual Control screen displays.
Use only Agilent-certified filters or Millipore-equivalent types of
filters. Other types of filters may not work properly in the filter
changer and could damage the unit.