Operating the 8000 4
8000 Operator’s Manual 103
Each unit starts running the program you entered.
Printer Operation and Communications
Use print options to
print batch information and sample times for each of the 15 stored
programs.
control the built-in Report Center Printer and the remote printer (if
one is connected).
set the communication port identification number and the baud rate.
Press PRINT. Enter a program number and press ENTER. The Print
Selections screen displays.
NOTE
Ensure no start output delays have been set on any of the 8000s.
Depending on your configuration, see “3 Start Output Delay (peristaltic
pump)” under “8000 Menu Options - Apparatus 1 / 2” on page 48 or “6
Start Output Delay” under “Main Menu Option 1 - Set
Clock/Alarms” on page 51.
NOTE
Ensure the printer type is set to impact for the instrument to operate properly
(MENU > 0 > E under “Hidden Key Functions”
on page 41)
***PRINT SELECTIONS***
1 BATCH INFORMATION 2 SAMPLE TIMES
3 TURN ON (OFF) REP CENTER 4 SET COM PORT ID
5 SET BAUD RATE