90 8000 Operator’s Manual
4 Operating the 8000
Program Option 2 - Sample Time Points
From the Program Variables screen, select option 2, Sample Time
Points. The following screen displays:
7 Filter Chg (syringe
pump and filter changer)
If your system configuration includes the syringe pump and filter changer, select option
7, Filter Chg. Enter the desired number of samples each filter should process before
being discharged and press ENTER. The recommended value is 1. The Program
Variables screen displays.
Note: If your system configuration includes the syringe pump and filter changer, ensure
the syringe pump option is enabled. From the Ready screen, press MENU. Select
option 3, Set Syringe Pump. Select option 1, Enable. Also press MENU > 0 > P and
select option 1, Enable.
The replacement media option is not available with this system configuration. Ensure
the RM option for the syringe pump is disabled using the hidden key function MENU >
0 > R.
7 R-M (syringe pump) If your system configuration includes the syringe pump but not the filter changer, select
option 7, R-M. The Replacement Media Option screen displays. Select option 2,
Disable. The Program Variables screen displays.
Note: The replacement media option is not available for use with the Apparatus 3 / 7.
The replacement media option must be disabled or the instrument will not operate
properly.
Table 2 2 Program Variables Options (Apparatus 3 / 7)
Option Function
SAMPLE TIME POINT: 000:02
ENTER hhh:mm UP TO 999:59 HOURS
PROG #: 1 ROW: 0 VOLUME: 5ml
TIME:01/01/12 09:37:14 ELAPSE:000:00:00
NOTE
This screen accepts sampling times and saves them for execution when
the program is run. Time points are requested in order of execution
starting with collection row number 0 on the sample tray. The sample
time is entered as hours and minutes. Once you enter a time, the row
number increments and another sample time point is requested.