Operating the 8000 4
8000 Operator’s Manual 59
Manual Operation Option 6 - Syringe/Filter
From the Manual Operations screen, select option 6, Syringe/Filter. The
Operate Syringe/Filter Changer screen displays.
Following is a description of the Operate Syringe/Filter Changer screen
options:
5 Turn On (Off) R-M
Pump
Select option 5, Turn On (Off) R-M Pump, to toggle the replacement media
option off and on, if installed.
6 Syringe/Filter (syringe
pump only)
Use this option to manually control the syringe pump and filter changer as
applicable. See “Manual Operation Option 6 - Syringe/Filter” below.
Table 8 Manual Operations Options (Apparatus 1 / 2)
Option Function
***OPERATE SYRINGE/FILTER CHANGER***
1 VALVE TO INPUT 2 VALVE TO OUTPUT
3 SYRINGE FILL 4 SYRINGE DISPENSE
5 FILTER CHANGER 6 SYRINGE PARAMETERS
Table 9 Operate Syringe/Filter Changer Options (Apparatus 1 / 2)
Option Function
1 Valve to Input The syringe pump contains a valve to control the direction of the flow of media.
Use this option to move the valve to the input position to manually input or fill the
syringe with media.
Select option 1, Valve to Input, to move the valve to the input position. The valve
emits a loud click when it moves. If you hear no sound, the valve is already in the
input position.
2 Valve to Output The syringe pump contains a valve to control the direction of the flow of media.
Use this option to move the valve to the output position to manually output or
purge the syringe of media.
Select option 2, Valve to Output, to move the valve to the output position. The
valve emits a loud click when it moves. If you hear no sound, the valve is already
in the output position.