the 8000 solenoid valves. Ensure each line is attached to its identical
color.
12 Connect the remaining length of tubing from the top positions of the
8000 solenoid valves to the return lines on the dissolution apparatus.
Ensure each line is attached to its identical color.
13 If the system configuration includes the 705-DS, connect the return
lines to the return cannulas of the resident probes.
Unpacking / Setup of 8000, Syringe Pump, and Filter
Changer
Environmental Requirements for Filter Changer Installation
• Altitude max.2000m;
• Humidity max relative humidity 80% for temperatures up to 31 ºC
decreasing linearly to 50% relative humidity at 40ºC;
• Indoor use only
• Temperature: 5 ºC to 40 ºC;
• Mains supply voltage fluctuations are not to exceed ± 10% of the
nominal supply voltage.
• Operating conditions 8 mins ON Max; 5 mins OFF min
Do not tighten the plastic fittings on the ends of the tubing more
than half a turn past finger-tight. Overtightening will damage
the fittings.