Setting Up the 8000 3
8000 Operator’s Manual 23
5 Loosely fit the peristaltic pump tubing along the inside groove of the
first cartridge, locking the tubing clips on the outside notches.
6 Refit the cartridge to the pump by hooking the clip on the bottom left
of the cartridge into the groove on the left side of the pump.
7 Push down the right side of the cartridge, ensuring that the tubing
remains under the cartridge, and pull up the lever to lock the
cartridge in place. If desired, trim the excess tubing to within one
inch of the tubing clips.
8 Repeat Steps 5 - 7 for each remaining cartridge.
9 Refit the cover and replace the two screws. If the cover is not
replaced correctly, the pump will not operate.
10 Refer to the following example diagram to correctly make cable
connections for the 8000, peristaltic pump, and dissolution
Ensure that all pump cartridges are installed. It is not necessary to install
tubing on any unused positions.
The cover does not fit unless the cartridges are positioned with the
levers on the right side.
The electrical connection at the back of the 8000 is the primary
disconnect for the instrument. The instrument should be
positioned to allow accessibility to the power cords for easy
disconnection.