3 Setting Up the 8000
MENU > 0 > D Use this key sequence to enable or disable the multiple sample option. The asterisk (*)
displays next to the current selection.
Note: For use with APP 3/7 (Apparatus 3 / 7) systems only.
MENU > 0 > E Use this key sequence to select the printer type. Ensure the asterisk (*) displays next
to option 1, Impact. If not, select option 1 to enable the impact printer.
Note: Select option 1, Impact, or the instrument will not operate properly.
MENU > 0 > F Use this key sequence to select the date display style. The asterisk (*) displays next to
the current selection. Select option 1, MM/DD/YY, or option 2, DD/MM/YY.
MENU > 0 > P Use this key sequence to enable or disable the syringe pump. The asterisk (*) displays
next to the current selection.
If your system configuration includes the syringe pump, you must enable the syringe
pump control options on the 8000 prior to setting the syringe pump parameters (see “3
SET SYRINGE PUMP” on page 48.
MENU > 0 > R Use this key sequence to enable or disable media replacement for the syringe pump.
The asterisk (*) displays next to the current selection. Select option 1, Enable, or
option 2, Disable, as appropriate.
Note: This option is not available for APP 3/7 (Apparatus 3 / 7) systems.