60 8000 Operator’s Manual
4 Operating the 8000
3 Syringe Fill Use this option to draw a specific amount of medium into the syringes. Ensure the
valve is in the input position before drawing the medium (“1 Valve to Input” on
page 59).
Select option 3, Syringe Fill. Enter the desired percentage of each syringe to fill
and press ENTER. The syringes draw the specified amount of medium.
4 Syringe Dispense Use this option to dispense medium from the syringes.
Select option 4, Syringe Dispense. The syringes dispense the previously drawn
medium (see “3 Syringe Fill” above).
5 Filter Changer Use this option to manually operate the filter changer.
Select option 5, Filter Changer. The Filter Change Manual Control screen
displays.
To open the filter clamp, select option 1, Open Clamp.
To close the filter clamp, select option 2, Close Clamp.
To change the filters, select option 3, Change Filters. Enter the number of filters to
change and press ENTER. The specified number of filters are expelled and
replaced.
Press ESC to return to the Operate Syringe/Filter Changer screen.
6 Syringe Parameters Use this option to configure the syringe pump. Select option 6, Syringe
Parameters. The Set Syringe Pump Parameters screen displays. For a description
of the screen options, see “Main Menu Option 3 - Set Syringe Pump” on page 55.
Note: The syringe pump parameters are global. These parameters can be entered
under option 3 of the Main Menu or option 6 of the Operate Syringe/Filter
Changer screen.
Table 9 Operate Syringe/Filter Changer Options (Apparatus 1 / 2)
Option Function