Operating the 8000 4
8000 Operator’s Manual 95
Manual Sample
From the Ready screen, press MANUAL SAMPLE to collect a sample
from the dissolution apparatus on command without a program. These
options are only for manual sampling and have nothing to do with the
stored program parameters for purge volume, prime volume, and so on.
When your system configuration includes the peristaltic pump, the
following Manual Sampling screen displays:
When your system configuration includes the syringe pump with or
without the filter changer, the following Manual Sampling screen
displays:
Following is a description of the Manual Sampling screen options:
***MANUAL SAMPLING***
1 START 2 SET ROW # 3 SET VOLUME
4 PRIME TIME 5 PURGE TIME 6 R-M OPTION
ROW 3 VOLUME 5ml PRIME 60 PURGE 60
***MANUAL SAMPLING***
1 START 2 SET ROW # 3 SET VOLUME
4 PRIME VOL. 5 PURGE VOL. 6 R-M OPTION
ROW 3 VOLUME 5ml PRIME 10 PURGE 20
Table 2 5 Manual Sampling Options
Option Function
1 Start Select option 1, Start, to start the manual sampling procedure using the values you
enter for options 2 through 6 on the Manual Sampling screen. The display screen
details each step of the manual collection as it takes place.
Note: Select this option last.
2 Set Row # Select option 2, Set Row #. Enter the row number between 0 and 9 where you want
the samples deposited in the sample tray and press ENTER. The Manual Sampling
screen displays.