Agilent Technologies ENA Series Guía de usuario Pagina 121

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 775
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 120
Chapter 4 115
Calibration
Enhanced Response Calibration
4. Calibration
Enhanced Response Calibration
In enhanced response calibration, calibration data are measured by connecting an OPEN
standard, a SHORT standard, or a LOAD standard to the output port (or a THRU standard
between two ports).This calibration effectively eliminates the directivity error, crosstalk,
source match error, frequency response reflection tracking error, and frequency response
transmission tracking error from the test setup in a transmission or reflection test using
those ports (Figure 4-13).
Figure 4-13 2-Port Error Model (Enhanced Response)
Vista de pagina 120
1 2 ... 116 117 118 119 120 121 122 123 124 125 126 ... 774 775

Comentarios a estos manuales

Sin comentarios