Chapter 15 531
Measurement Examples
Evaluating Transmission Characteristics of a Front End Module
15. Measurement Examples
10. Executing Measurement
Step 1. Display the trigger menu.
Step 2. Set the trigger source to “manual.”
Step 3. Set the trigger mode for channel 1 to “continuous.”
Step 4. Set the trigger mode for channels 2, 3, and 4 to “continuous” according to the same
procedure as in Step 3.
Step 5. Execute the measurement.
Step 6. Repeat the following procedure to execute auto scale for all of the channels.
Step 7. After all of the procedures are completed, the screen shown in Figure 15-38 appears.
Description Key Operation
Display the trigger menu.
Setting Description Key Operation
Set the trigger source to “manual.”
Trigger Source - Manual
Setting Description Key Operation
Set channel 1 to the active channel.
Set the trigger mode to “continuous.”
Continuous
Setting Description Key Operation
Generate a trigger event.
Trigger
Setting Description Key Operation
Set the active channel.
Execute auto scale.
- Auto Scale