666 Appendix D
Softkey Functions
Average Menu
Average Menu
Key Operation Function SCPI Command
Displays softkeys for setting averaging options.
Averaging Restart Resets the counter and restarts from “1”.
:SENS{1-16}:AVER:CLE
Avg Factor
Sets the averaging factor. The averaging factor must be defined as an
integer from 1 to 999.
:SENS{1-16}:AVER:COUN
Averaging
Enables or disables averaging execution.
:SENS{1-16}:AVER
Ave Trigger
Enables or disables averaging trigger.
:TRIG:AVER
Smo Aperture
Sets the smoothing aperture using a percentage against the sweep span
:CALC{1-16}:SMO:APER
Smoothing
Enables or disables the smoothing function. When enabled, the letters
“Smo” will appear in the trace status area.
:CALC{1-16}:SMO:STAT
IF Bandwidth
Sets the IF bandwidth. For more about the effective range of the IF
bandwidth, refer to “Specifications and Reference Data”. If an input value
is out of the range, it will automatically be modified to the closest value in
the effective range. A narrow IF bandwidth slows down the sweep speed,
but improves the S/N ratio.
:SENS{1-16}:BAND or
:SENS{1-16}:BWID
Return
Returns to the “E5070B/E5071B Menu (Top Menu)” on page 650.