542 Chapter 16
Specifications and Supplemental Information
Corrected System Performance
Table 16-4 Corrected System Performance With 3.5 mm Device Connector Type, 85033E
Calibration Kit
Network analyzer: E5070B/E5071B, Calibration kit: 85033E (3.5 mm, 50 Ω), Calibration: full 2-port
IF bandwidth = 10 Hz, No averaging applied to data, Environmental temperature = 23°C±5°C with <1°C deviation
from calibration temperature, Isolation calibration not omitted
Description
Specification (dB)
3 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz
Directivity 46 38 38
Source Match 43 37 36
Load Match 46 38 38
Reflection Tracking ±0.006 ±0.009 ±0.010
Transmission Tracking ±0.016 ±0.065 ±0.079
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)