Appendix A 597
Manual Changes
Manual Changes
A. Manual Changes
Step 1. Measure calibration data.
In this example, 3-port TRL/LRM calibration for test ports 1, 2, and 3 (1 in Figure A-8) is
performed by using the Match, Line 1, and Line 2 standards.
Figure A-8 Performing multiport TRL/LRM calibration
Step 2. Press the
Thru button (2 in Figure A-8).
Step 3. Make the through condition for test ports 1 and 2 and press the
[ ]Thru 1-2 button (1 in
Figure A-9). The through measurement is performed, and an asterisk appears at the
[ ]Thru
1-2
button.
Step 4. Make the through condition for test ports 1 and 3, and press the
[ ]Thru 1-3 button (2 in
Figure A-9). The through measurement is performed, and an asterisk appears at the
[ ]Thru
1-3
button and then a check mark [v] (3 in Figure A-9) also appears on the Thru button.
Calibration Type Measurement ports of
Reflection and Match cal.
Measurement paths of Thru
and Line cal.
3-port TRL/LRM cal. for
Ports: 1-2-3
All ports (1, 2, and 3) 1-2 and 1-3
3-port TRL/LRM cal. for
Ports: 1-2-4
All ports (1, 2, and 4) 1-2 and 1-4
3-port TRL/LRM cal. for
Ports: 1-3-4
All ports (1, 3, and 4) 1-3 and 3-4
3-port TRL/LRM cal. for
Ports: 2-3-4
All ports (2, 3, and 4) 2-3 and 3-4
4-port TRL/LRM cal. for
Ports: 1-2-3-4
All ports (1, 2, 3, and 4) 1-2, 1-3, and 3-4