Chapter 16 549
Specifications and Supplemental Information
Test Port Input
16. Specifications and
Supplemental Information
Table 16-12 Test Port Input (Stability)
Description Specification Supplemental Information
Stability Magnitude
*1
3 MHz to 3 GHz 0.005 dB/°C
(at 23 °C±5°C, typical)
3 GHz to 6 GHz 0.01 dB/°C
(at 23 °C±5°C, typical)
6 GHz to 8.5 GHz 0.04 dB/°C
(at 23 °C±5°C, typical)
Stability Phase
*1
3 MHz to 3 GHz 0.1 °/°C
(at 23 °C±5°C, typical)
3 GHz to 6 GHz 0.2 °/°C
(at 23 °C±5°C, typical)
6 GHz to 8.5 GHz 0.8 °/°C
(at 23 °C±5°C, typical)
*1.Stability is defined as a ratio measurement at the test port.
Table 16-13 Test Port Input (Dynamic Accuracy)
Accuracy of the test port input power reading is relative to 10 dBm reference input power level.
Specification Supplemental Information