162 Chapter 4
Calibration
Adapter Characterization
How to execute adapter characterization
1. Setting Stimulus Conditions
Set the stimulus conditions of the channel for which you will execute the adapter
characterization.
For information on the setting procedure, see Chapter 3, “Setting Measurement
Conditions.”
2. Performing Calibration
Perform 1-port calibration for test port 1 and test port 2 in the channel for which the
stimulus condition has been set. Use the connector type appropriate for the adapter for the
calibration surface of test port 1 and test port 2.
For information on the procedure for performing this calibration, see “1-Port Calibration
(reflection test)” on page 113.
3. Starting VBA Macro
Step 1. Press .
Step 2. Press
Load Project.
Step 3. The Open dialog box appears. Specify the file name
“D:\Agilent\AdapterCharacterization.vba” and press the
Open button.
Step 4. Press to start the macro.
Step 5. Select
Characterize Adapter (1 in Figure 4-41) to display the Adapter Characterization
screen.
Figure 4-41 Adapter Characterization macro (Adapter Characterization screen)
4. Selecting Channel
Select the channel for which calibration has been performed (2 in Figure 4-41).
NOTE The selected channel has no relation to the active channel.