Vista de pagina 180
Chapter 4 175
Calibration
User-characterized ECal
4. Calibration
Step 4. The dialog box shown in Figure 4-50 is displayed to confirm execution. Press OK (1 in
Figure 4-50) to start storing the user characteristics.
CAUTION Do not disconnect the USB cable or terminate the VBA macro by force while the VBA
macro is storing data to the ECal's built-in flash memory. Doing so may damage the
ECal module.
Figure 4-50 Execution Confirmation screen
The dialog box shown in Figure 4-51 appears while the VBA macro is storing data to
memory. Storing the user characteristics takes a few minutes depending on the amount of
data.
Figure 4-51 Screen while macro is storing data
Step 5. Another dialog box is displayed to notify completion of data storage. Click
OK.
8. Closing the VBA macro
Press
Close (3 in Figure 4-47) to close the macro.
Vista de pagina 180