Chapter 12 411
Optimizing Measurements
Reducing Measurement Error in High Temperature Environments
12. Optimizing Measurements
Reducing Measurement Error in High Temperature
Environments
The E5070B/E5071B is designed to obtain the best measurement accuracy at the ambient
temperature range of 23°C ± 5°C. The high temperature measurement mode of the
E5070B/E5071B reduces measurement error (drift error) at an ambient temperature of
28°C to 33°C.
NOTE The high temperature measurement mode must be turned off when the analyzer is used at
an ambient temperature below 28°C. Otherwise, the measurement accuracy may be
degraded.
Procedure
Step 1. Press .
Step 2. Press
Service Menu.
Step 3. Press
High Temperature to turn ON/OFF the high temperature measurement mode.