232 Chapter 4
Calibration
Vector-Mixer Calibration
the macro (1 in Figure 4-73).
NOTE If failure occurs when reading the data file for the calibration mixer with IF filter, the
characterization may have done by using only one port instead of using two ports.
Figure 4-73 Balanced Mixer Characterization Macro
Step 9. Select the measurement port as shown in Figure 4-73 (2 in Figure 4-73) and then connect a
THRU between the IF ports of the calibration mixers to correct the phase difference of the
LO signals for the calibration mixers with IF filters.
Step 10. Pressing the
Execute (3 in Figure 4-73) button executes a phase error correction and
overwrites the results on the original data file.
NOTE You cannot run the Execute function when selecting the measurement port if the data file
of the calibration mixer with IF filter (*.s2p) has not been set for the network removal
function of the fixture simulator.
NOTE The phase error correction data reflects the phase difference of the LO signals for the phase
information of the calibration mixers data file, which is registered in any two ports.
Step 11. Press the
Close button (4 in Figure 4-73) to exit macro.