Appendix A 593
Manual Changes
Manual Changes
A. Manual Changes
NOTE When you define each calibration standard, the following points should be considered.
Reference impedance Z0 must be set to the same value as the ENAs system impedance
Z0 value.
When you use Line 1, Line 2, and Line 3, their defined frequency ranges must overlap
by at least 10 kHz.
Saving and loading calibration kit definitions
Press the
Save button (11 in Figure A-4) to save the definition of the current calibration kit
to a file. Press the
Recall button (12 in Figure A-4) to recall the definition of a calibration
kit from a previously saved file.
NOTE If you save as “D:\Agilent\Data\TRL_LRM_cal\Default.dat,” the file is handled as the
default definition file. The default definition file is automatically recalled when the macro
starts.
NOTE The factory-shipped default definition file has the same content as
“D:\Agilent\Trldata\SysDefault.dat.” Copy “D:\Agilent\Trldata\SysDefault.dat” to the
default definition file in order to restore the default definition file to its factory-shipped
condition.
Do not change “D:\Agilent\Trldata\SysDefault.dat.”
Initializing calibration kit definition
Press the
default button (13 in Figure A-4) to recall the definition of a calibration kit from
the default definition file (“D:
\Agilent\Data\TRL_LRM_cal\Default.dat”).
Step 3. Press the
Close button (14 in Figure A-4) to finish defining the calibration kit.
3. Performing Calibration
Measure necessary calibration data and enable error correction.
NOTE The definition of the frequency range of the line standard and match standard used for
measurement must cover the sweep range of the channel for which you perform
calibration.
If the
Thru, Reflection, Match, and Line keys do not appear, go back to the calibration kit
definition menu and confirm that the frequency range definitions of match and lines are
correctly entered to cover the ENAs measurement frequency range.
Step 1. Measure calibration data.
In this example, 2-port TRL/LRM calibration between test ports 1 and 2 (1 in Figure A-5)
is performed by using the Match, Line 1, and Line 2 standards.
Vista de pagina 598