510 Chapter 15
Measurement Examples
Measuring the Deviation from a Linear Phase
2. Define the Measurement Conditions
The measurement conditions are defined by following the steps described below.
3. Execute the Calibration
The THRU response calibration is executed.
4. Connect the DUT and Execute the Auto Scale
The DUT is connected again as shown in Figure 15-21 to execute the auto scale.
Setup Description Key Operation
Presetting
-
OK
Center frequency: 1.09 GHz
Frequency span: 20 MHz
Measurement parameter: S
21
- S21
Data format: Expand Phase
-
Expand Phase
Executing the Auto Scale
-
Auto Scale
Setup Description Key Operation
Executing the THRU response (A THRU standard is connected instead of a
DUT) -
Calibrate - Response (Thru) -
Thru - Done
Setup Description Key Operation
Executing the auto scale
-
Auto Scale