Chapter 15 519
Measurement Examples
Measuring an Unbalanced and Balanced Bandpass Filter
15. Measurement Examples
6. Extending the Calibration Plane (removing the cause of error)
In this section you will use the port extension function to remove an electrical delay caused
by cables or fixtures located between the calibration reference plane and the DUT to be
evaluated. If you can provide a two-port Touchstone data file representing the
characteristics of the network to be removed, the network removal function allows you to
remove the network and extend the calibration reference plane.
Follow the procedure below to set port extension for each test port.
Figure 15-29 shows the results of extending the calibration reference plane.
Figure 15-29 Results of extending the calibration reference plane
7. Setting the Port Reference Impedances
With the reference impedances of two test ports in unbalanced measurements set to Z
0
,
conversion of those ports into balanced ports permits the impedance of the balanced ports'
common mode to be automatically set to Z
0/2
and the impedance of their differential mode
to be automatically set to 2Z
0
.
Setting Description Key Operation
Port extension of test port 1: 260 ps
-
Port Extensions - Extension Port 1 -
Port extension of test port 2: 260 ps
Extension Port 2 -
Port extension of test port 3: 260 ps
Extension Port 3 -
Port extension: ON
Extensions (turns it ON)