560 Chapter 16
Specifications and Supplemental Information
Measurement Throughput Summary
Table 16-22 Typical Cycle Time for Measurement Completion
*1*2
(ms)
Number of Points
51 201 401 1601
Start 1 GHz, Stop 1.2 GHz, 100 kHz IF bandwidth
Uncorrected 7 17 29 90
2-port cal 123255178
Start 300 kHz, Stop 3 GHz, 100 kHz IF bandwidth
Uncorrected 14 27 43 130
2-port cal 265084258
Start 300 kHz, Stop 8.5 GHz, 100 kHz IF bandwidth
Uncorrected 16 30 49 146
2-port cal 305796291
*1.Typical performance.
*2.Sweep mode: Std Stepped. Analyzer display turned off with :DISP:ENAB OFF. Number of traces = 1. System
error correction: ON
Table 16-23 Cycle Time (ms)
*1*2
vs. Number of Points
Number of
Points
Sweep mode: Fast Swept
System error correction: OFF
Sweep mode: Fast Swept
System error correction: ON
Sweep mode: Std Stepped
System error correction: ON
3444
11444
51447
101 4 5 11
201 5 6 17
401 8 8 29
801 11 13 52
1601 18 23 90
*1.Typical performance.
*2.Start 1 GHz, Stop 1.2 GHz, 100 kHz IF bandwidth, Error correction: OFF, Display update: OFF, Number of traces