Chapter 4 209
Calibration
Power Calibration
4. Calibration
Preparing power meter and sensor
To execute power calibration, you need to prepare the power meter and power sensor used
to acquire power calibration data. Table 4-9 shows available power meters and
recommended power sensors for power calibration.
Table 4-9 Available power meters and recommended power sensors
Model name
Power meters Agilent 437B
Agilent 438A
Agilent E4416A
Agilent E4417A
Agilent E4418A
Agilent E4418B
Agilent E4419A
Agilent E4419B
Agilent N1911A
Agilent N1912A
HP EPM-441A
HP EPM-442A
Power sensors Agilent 8482A
Agilent E4412A
HP ECP-E18A