Agilent Technologies ENA Series Guía de usuario Pagina 356

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 775
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 355
350 Chapter 9
Analysis in Time Domain (Option 010)
Deleting Unnecessary Data in Time Domain (gating)
range.
NOTE You can set the center and span by dragging and dropping flags indicating the gate range
(Figure 9-10).
Enabling gating function
When you enable the gating function, data within the specified range is deleted. When the
transformation function is enabled, the flags indicating the gate range is displayed as
shown in Figure 9-10.
NOTE In Figure 9-10, the gate type is set to band pass. When it is set to notch, the directions of the
flags indicating the ends of the gate range are reversed.
Figure 9-10 Screen when transformation function and gate function are enabled
Operational procedure
Step 1. Press (or ) and (or ) to activate the trace for
which you want to use the gate function.
Step 2. - Press
Gating to display the “Gating” menu.
Step 3. Use
Gating to enable (ON) the gate function.
Vista de pagina 355
1 2 ... 351 352 353 354 355 356 357 358 359 360 361 ... 774 775

Comentarios a estos manuales

Sin comentarios