142 Chapter 4
Calibration
3-port TRL calibration
coefficient (by pressing Done).
Step 7.
Thru/Line to start the measurement of the calibration standard.
Step 8.
Reflect to start the measurement of the calibration standard.
Step 9.
Line/Match to start the measurement of the calibration standard.
Softkey Function
Port x-y Thru Executes thru/line calibration for test ports x and y.
Port x-z Thru Executes thru/line calibration for test ports x and z.
Port y-z Thru Executes thru/line calibration for test ports y and z.
Softkey Function
Port x Reflect Executes reflection calibration for test port x.
Port y Reflect Executes reflection calibration for test port y.
Port z Reflect Executes reflection calibration for test port z.
Softkey Function
x-y Line/Match
*1
*1.This is equivalent to measuring x-y Fwd (Syx) and x-y Rvs (Sxy).
Executes line/match calibration for test ports x and y. Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
x-y Fwd (Syx)
*2
*2.Measuring x-y Fwd (Syx) and x-y Rvs (Sxy) is equivalent to measuring x-y Line/Match.
Executes line/match calibration for test ports x and y. Use this to perform
forward-direction measurement only.
x-y Rvs (Sxy)
*2
Executes line/match calibration for test ports x and y. Use this to perform
reverse-direction measurement only.
x-z Line/Match Executes line/match calibration for test ports x and z. Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
x-z Fwd (Szx) Executes line/match calibration for test ports x and z. Use this to perform
forward-direction measurement only.
x-z Rvs (Sxz) Executes line/match calibration for test ports x and z. Use this to perform
reverse-direction measurement only.
y-z Line/Match Executes line/match calibration for test ports y and z. Use this to perform
forward-direction measurement and reverse-direction measurement at the
same time.
y-z Fwd (Szy) Executes line/match calibration for test ports y and z. Use this to perform
forward-direction measurement only.
y-z Rvs (Syz) Executes line/match calibration for test ports y and z. Use this to perform
reverse-direction measurement only.