122 Chapter 4
Calibration
Full 3-Port Calibration
Figure 4-18 Connecting standards in full 3-port calibration
Procedure
Step 1. Press or to select the channel for which you want to perform the
calibration.
Step 2. Press .
Step 3. Press
Calibrate.
Step 4. Press
3-Port Cal.
Step 5. Press
Select Ports.
Step 6. Select the test ports on which you will perform full 3-port calibration. (In the procedure
below, the selected test ports are denoted as x, y, and z.)
NOTE An asterisk (*) in the upper-right of the softkey indicates that the existing calibration
coefficient will be cleared if you select the test port and execute acquisition of the
Softkey Function
1-2-3 Selects test ports 1, 2, and 3
1-2-4 Selects test ports 1, 2, and 4
1-3-4 Selects test ports 1, 3, and 4
2-3-4 Selects test ports 2, 3, and 4