166 Chapter 4
Calibration
Adapter Characterization
6. Measuring Data
Step 1. Connect the probe to the end of the test fixture on the DUT side.
Step 2. Depending on the setting of Cal Kit 1, connect each standard to the connector side of the
test fixture and then press the corresponding button (4 in Figure 4-41). When the data
measurement is complete, the button turns yellow.
7. Saving to file
Press the
Save button (6 in Figure 4-41) to save the calculated S-parameters to a 2-port
Touchstone file.
8. Closing VBA Macro
Press
Close (7 in Figure 4-41) to close the macro.