Appendix D 673
Softkey Functions
Calibration Menu
D. Softkey Functions
(Continued)
Calibrate
4-Port Cal
*2
Transmission
Port 2-3 Thru
Executes a THRU calibration between test ports 2 and 3. :SENS{1-16}:CORR:COLL:THRU 2,3
:SENS{1-16}:CORR:COLL:THRU 3,2
Port 2-4 Thru
Executes a THRU calibration between test ports 2 and 4. :SENS{1-16}:CORR:COLL:THRU 2,4
:SENS{1-16}:CORR:COLL:THRU 4,2
Port 3-4 Thru
Executes a THRU calibration between test ports 3 and 4. :SENS{1-16}:CORR:COLL:THRU 3,4
:SENS{1-16}:CORR:COLL:THRU 4,3
Return
Returns to the softkey display screen one level higher.
Isolation
(Optional)
Displays softkeys for executing isolation calibrations.
Port 1-2 Isol Executes an isolation calibration on test ports 1 and 2. :SENS{1-16}:CORR:COLL:ISOL 1,2
:SENS{1-16}:CORR:COLL:ISOL 2,1
Port 1-3 Isol
Executes an isolation calibration on test ports 1 and 3. :SENS{1-16}:CORR:COLL:ISOL 1,3
:SENS{1-16}:CORR:COLL:ISOL 3,1
Port 1-4 Isol
Executes an isolation calibration on test ports 1 and 4. :SENS{1-16}:CORR:COLL:ISOL 1,4
:SENS{1-16}:CORR:COLL:ISOL 4,1
Port 2-3 Isol
Executes an isolation calibration on test ports 2 and 3. :SENS{1-16}:CORR:COLL:ISOL 2,3
:SENS{1-16}:CORR:COLL:ISOL 3,2
Port 2-4 Isol
Executes an isolation calibration on test ports 2 and 4. :SENS{1-16}:CORR:COLL:ISOL 2,4
:SENS{1-16}:CORR:COLL:ISOL 4,2
Port 3-4 Isol
Executes an isolation calibration on test ports 3 and 4. :SENS{1-16}:CORR:COLL:ISOL 3,4
:SENS{1-16}:CORR:COLL:ISOL 4,3
Return
Returns to the softkey display screen one level higher.
Done Terminates a calibration process and calculates the calibration
coefficients.
:SENS{1-16}:CORR:COLL:SAVE
Overwrite
Terminates partial overwrite and recalculates the calibration
coefficients.
:SENS{1-16}:CORR:COLL:PART:SAVE
Cancel
Displays softkeys for canceling a calibration.
OK Cancels a calibration process and clears the measurement data of
standard.
:SENS{1-16}:CORR:COLL:CLE
Cancel Returns to the softkey display screen one level higher.
Return Returns to the softkey display screen one level higher.
Return Returns to the softkey display screen one level higher.
2-Port TRL Cal Displays softkeys for executing 2-port TRL calibrations.
Select Ports
*1
Displays softkeys for selecting test ports.
1-2 Selects test ports 1 and 2.
:SENS{1-16}:CORR:COLL:METH:TRL2 1,2
1-3
Selects test ports 1 and 3.
:SENS{1-16}:CORR:COLL:METH:TRL2 1,3
1-4
*2
Selects test ports 1 and 4.
:SENS{1-16}:CORR:COLL:METH:TRL2 1,4
2-3
Selects test ports 2 and 3.
:SENS{1-16}:CORR:COLL:METH:TRL2 2,3
2-4
*2
Selects test ports 2 and 4.
:SENS{1-16}:CORR:COLL:METH:TRL2 2,4
3-4
*2
Selects test ports 3 and 4.
:SENS{1-16}:CORR:COLL:METH:TRL2 3,4
Cancel
Returns to the softkey display screen one level higher.
Key Operation Function SCPI Command
Vista de pagina 678