116 Chapter 4
Calibration
Enhanced Response Calibration
Figure 4-14 Connecting the Standard at Enhanced Response Calibration
Procedure
Step 1. Press or to select the channel for which you want to perform the
calibration.
Step 2. Press .
Step 3. Press
Calibrate.
Step 4. Press
Enhanced Response.
Step 5. Press
Ports to select the test ports on which an enhanced response calibration will be
performed.
Softkey display Function
2-1(S21 S11) Measure between test ports 1 and 2. selecting port 1 as the output.
3-1(S31 S11) Measure between test ports 1 and 3. selecting port 1 as the output.
4-1(S41 S11) Measure between test ports 1 and 4. selecting port 1 as the output.
1-2(S12 S22) Measure between test ports 2 and 1. selecting port 2 as the output.
3-2(S32 S22) Measure between test ports 2 and 3. selecting port 2 as the output.
4-2(S42 S22) Measure between test ports 2 and 4. selecting port 2 as the output.
1-3(S13 S33) Measure between test ports 3 and 1. selecting port 3 as the output.
2-3(S23 S33) Measure between test ports 3 and 2. selecting port 3 as the output.
4-3(S43 S33) Measure between test ports 3 and 4. selecting port 3 as the output.