Chapter 13 467
Setting and Using the Control and Management Functions
System Recovery
13.Setting and Using the
Control and Management
Functions
Step 8. The message as shown below appears. Press of the keyboard. If you want to cancel the
factory recovery, press here.
You chose to Restore your system by installing the original factory
installed OS and system software.
WARNING: Press C to Continue only if you are sure that you want
to proceed. The C: Drive will be completely overwritten with no
chance of recovering any data. Use Option 1 to recover the system
from a serious malfunction caused by corrupted or inadvertently
deleted files on the system's primary C: partition.
Press C to Continue or E to Exit: _
Step 9. The message as shown below appears. Press of the keyboard to start the factory
recovery. If you want to cancel the factory recovery, press here.
CAUTION! Interrupting this process may leave the system in an
unstable state. Allow the software to complete the
backup and recovery process. This may take up to 20
minutes depending on the system configuration.
Press C to Continue or E to Exit: _
CAUTION Never turn off the power during the system recovery because doing so may cause
serious damage to the E5070B/E5071B.
Step 10. The factory recovery will be complete in about 5 minutes. When the factory recovery is
complete, the message as shown below appears. Press , , and of the
keyboard at the same time to restart.
Remove the disk and Press CLT+ALT+DEL to restart your
system.
NOTE If the above message does not appear, the instrument is at fault; contact your local Agilent
customer center listed at the end of this manual or distributor.
Step 11. After restart, the screen for initial registration appears. Execute initial registration. For
information on the execution procedure, refer to
Installation/Quick Start Guide
.
Step 12. For the E5070B/E5071B equipped with the Option 016 touch screen, execute the
calibration of the touch screen. For information on the execution procedure, refer to
“Calibration of the Touch Screen” on page 474.