2-26 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
Measurement Modes
For 0 V output, enter the DZ command that is used to memorize the present
settings of the channel and change the output to 0 V.
The index data (max. 9999999) and the time data returned with the measurement
data will be as shown in the following formula. However, long measurement or busy
status may cause unexpected time and index data.
time = t + Bias hold time + ( index -1) × Sampling interval
Where, t is the time of the sampling measurement time origin, and is the time when
the output value is changed from base to bias.
NOTE The MT command sets the bias hold time, sampling interval, number of samples,
and base hold time. If the sampling interval is less than 2 ms, the SPGU cannot be
used.
The MSC command sets the automatic abort function and the post measurement
output value of SMU.
The ML command sets the linear sampling mode or the log sampling mode.
The MV and MI commands set the SMU synchronous voltage and current source
respectively. The base and bias values must have the same polarity for MI output.
The SIM 0 command sets the pulse generator operation mode for all SPGU
channels.
The SPRM command sets the output operation mode for all SPGU channels, free
run (pulse output continues until SPP), pulse count, or duration.
The SPPER command sets the pulse period for all SPGU channels.
The SPM command sets the output mode of the SPGU channel, DC voltage, 2-level
pulse using source 1, 2-level pulse using source 2, or 3-level pulse using sources 1
and 2.
The SPT command sets the delay time, pulse width, leading time, and trailing time
of the SPGU channel.
The SPV command sets the pulse base and peak voltage or the DC bias voltage of
the SPGU channel.
The MSP command sets the SPGU synchronous voltage source and its post
measurement output value.
The MCC command clears the MV, MI, and MSP command setting.
To Stop Output An automatic abort function is available. Refer to “Automatic Abort Function” on
page 2-46.