4-236 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
WZ?
WZ?
This query command immediately confirms all channel output, and returns the
status 0 if it is within ± 2 V or 1 if it is more than ± 2 V.
Syntax WZ? [timeout]
Parameters timeout : Timeout. Numeric expression.
0 to 655.35 sec, with 0.01 sec resolution.
With timeout parameter, this command waits until all channel output
becomes within ± 2 V or until the specified timeout elapses, and returns
0 or 1.
The WZ? 0 command has the same effect as the WZ? command.
Query Response state<CR/LF^EOI>
0: All channel output is within ± 2 V.
1: Any output channel applies more than ± 2 V.
Example
Statement
OUTPUT @B1500;"WZ? 5.0"
ENTER @B1500;A
XE
The XE command triggers the B1500 to start measurement, or causes the B1500 to
recover from the wait state set by the PA command. This command is not available
for the high-speed spot measurement.
Syntax XE
Example
Statement
OUTPUT @B1500;"XE"
NOTE After measurement, the measurement data will be entered to the output data buffer.
For data output format, refer to “Data Output Format” on page 1-25.
Execution
Conditions
The following execution conditions are for you who use the XE command to start
measurement. There is no execution condition when you use the XE command to
recover from the wait state.