Agilent Technologies B1500A Manual de servicio Pagina 45

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 598
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 44
Agilent B1500A/B1505A Programming Guide, Edition 11 1-27
Programming Basics
Data Output Format
Data Format
The data output format depends on the measurement mode as shown below.
High speed spot Data <terminator> (by TI, TV, TMACV, or TMDCV command)
Time,Data <terminator> (by TTI or TTV command)
Para1,Para2 <terminator> (by TIV or TC command)
Time,Para1,Para2 <terminator> (by TTIV or TTC command)
Data is the value measured by the channel you specify in the command. Time is the
time from timer reset to the start of measurement. Para1 and Para2 are current (A)
and voltage (V) respectively measured by the SMU or the primary and secondary
parameters (ex: Cp and G) respectively measured by the CMU. The CMU
measurement data is selected by the IMP command. See Table 4-16 on page 4-29.
TDI, TDV, TSQ,
TACV, TDCV
command
Time <terminator>
Time is the time from timer reset to the start of output.
Spot,
Multi channel
pulsed spot
Data1 [,Data2] . . . . <terminator>
DataN (N: integer) is the value measured by a channel. The order of Data is defined
by the MM command.
Pulsed spot,
Quasi-pulsed spot,
Spot C,
Pulsed spot C
Data <terminator>
Data is the value measured by the channel you specify by using the MM command.
For the spot C measurement, Data will be Para1,Para2 [,Osc_level,Dc_bias].
For the pulsed spot C measurement, Data will be Para1,Para2.
Para1 and Para2 are the primary and secondary parameters (ex: Cp and G). They
are selected by the IMP command. See Table 4-16 on page 4-29. And Osc_level and
Dc_bias are the monitor values of the oscillator level (AC signal level) and the DC
bias respectively. They are sent if the data output is enabled by the LMN command.
Linear search,
Binary search
Data_search [,Data_sense]<terminator>
This is the data at the measurement point closest to the search target. Data_search is
the value forced by the search output channel. Data_sense
is the value measured by
the search monitor channel. It is sent if data output is enabled by the BSVM
command for the binary search, or the LSVM command for the linear search.
Vista de pagina 44