Agilent B1500A/B1505A Programming Guide, Edition 11 4-175
Command Reference
SOPC?
If the automatic abort function for sweep source is enabled by the WM command,
the sweep measurement is automatically stopped by detecting “compliance”.
Syntax SOPC chnum,power
Parameters chnum : Channel number assigned to the measurement resource. Integer
expression. 1 to 10 or 101 to 1001. See Table 4-1.
power : Comparison reference value (in W). Numeric expression. Resolution:
0.001 W.
0.001 to 22500 for UHCU, 0.001 to 100 for UHVU (DC), 0.001 to 200
for UHVU (pulse), 0.001 to 600 for HVMCU
Example
Statements
OUTPUT @B1500;"SOPC 3,100"
SOPC?
This query command is effective for the HVMCU, UHCU, and UHVU. This query
command returns the comparison reference value set by the SOPC command.
Syntax SOPC? chnum
Parameters chnum : Channel number assigned to the measurement resource. Integer
expression. 1 to 10 or 101 to 1001. See Table 4-1.
Query Response power<CR/LF^EOI>
Comparison reference value (in W).
Example
Statements
OUTPUT @B1500;"SOPC? 3"
ENTER @B1500;A
SOVC
This command is effective for the HVMCU, UHCU, and UHVU. This command
enables the output voltage comparison function.
This function automatically measures the voltage applied to the device under test
and compares the measurement result with the voltage value specified in this
command. Then if it is not less than the voltage value, the measurement status is set
to “compliance”.
This function is disabled by setting 0 to voltage.