4-44 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
AZ
Number of samples = number
where initial number means the number of samples the Agilent B1500
automatically sets and you cannot change. For voltage measurement,
initial number=1. For current measurement, see Table 4-22.
If you select the manual mode, number must be initial number or more
to satisfy the specifications.
Table 4-22 Initial Number for Current Measurement
Example
Statements
OUTPUT @B1500;"AV 10"
OUTPUT @B1500;"AV -50"
OUTPUT @B1500;"AV 100,1"
AZ
This command is used to enable or disable the ADC zero function that is the
function to cancel offset of the high-resolution A/D converter. This function is
especially effective for low voltage measurements. Power on, *RST command, and
device clear disable the function.
This command is effective for the high-resolution A/D converter, not effective for
the high-speed A/D converter.
Syntax AZ mode
Parameters mode : Mode ON or OFF.
0: OFF. Disables the function. Initial setting.
1: ON. Enables the function.
Current Measurement Range
Voltage Output Range
a
a. For measurement channels that force current, this is the minimum range
that covers the voltage compliance value.
to 40 V 100 V 200 V
to 10 μA 4 10 25
100 μA to 1 A 111